Tengku Azmi, Tengku Sharifah Marliza and Mohd. Yusoff, Abdull Rahim and Abdul Karim, K. J. (2010) Determination of Vanadium(IV) and Vanadium(V) in benfield samples by IEC with conductivity detection. Chromatographia, 72 (1). 141 - 144. ISSN 0009-5893 (Print) , 1612-1112 (Online)
Full text not available from this repository.
Official URL: http://dx.doi.org/10.1365/s10337-010-1630-x
Abstract
In this study, the determination of vanadium valence state, V(IV) and V(V) has been achieved using ion-exchange chromatography with conductivity detector. In this method, V(IV) was determined as V(IV)-EDTA complex and V(V) as vanadate ion. Determination of V(IV) was successfully done using 3 mM carbonate/bicarbonate/EDTA at pH 8.6 as the eluent. The additive, EDTA in the mobile phase did not seem to interfere with the V(IV) analysis. The detection of V(V) was achieved with 5 mM disodium hydrogen phosphate buffer at pH 10.4. A linear calibration graph over VO3− and V(IV) with concentration ranges 5–15 mg L−1 gave the detection limit at 0.09 and 0.1 mg L−1, respectively. Both V(IV) and V(V) were successfully determined in Benfield sample, with concentrations of V(IV) and V(V) at 4 and 11,000 mg L−1, respectively.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | conductivity detection, benfield sample, vanadium |
Subjects: | T Technology > TA Engineering (General). Civil engineering (General) |
Divisions: | Civil Engineering |
ID Code: | 25097 |
Deposited By: | Narimah Nawil |
Deposited On: | 16 May 2017 10:28 |
Last Modified: | 21 Mar 2018 00:38 |