Zainudin, Nur Amalina and Ramlee, Muhammad Hanif and Mohd. Latip, Hadafi Fitri and Azaman, Aizreena and Gan, Hong Seng and Evelyn, Garcia-Nieto and Abdul Kadir, Mohammed Rafiq (2019) Biomechanical evaluation of pin placement of external fixator in treating tranverse tibia fracture: analysis on first and second cortex of cortical bone. Malaysian Journal of Fundamental and Applied Sciences, 15 (1). pp. 75-79. ISSN 2289-5981
|
PDF
1MB |
Official URL: https://dx.doi.org/10.11113/mjfas.v15n2019.1263
Abstract
Biomechanical perspective of external fixator is one of the greatest factor to consider in successfully treating bone fracture. This is due to the fact that mechanical behavior of the structure can be analyzed and optimized in order to avoid mechanical failure, increase bone fracture healing rate and prevent pre-term screw loosening. There are three significant factors that affect the stability of external fixator which are the placement of pin at the bone, configuration and components of external fixator. These factors lead to one question: what is the optimum pin placement in which exerts optimum stability? To date, literature on above mentioned factors is limited. Therefore, we conducted a study to evaluate the uniplanar-unilateral external fixator for two different pin placement techniques in treating transverse tibia fracture via finite element method. The study was started off with the development of transverse tibia fracture using Mimics software. Computed tomography (CT) data image was utilized to develop three dimensional tibia bone followed by crafting fracture on the bone. Meanwhile, the external fixator was developed using SolidWork software. Both tibia bone and external fixator were meshed in 3-matic software with triangular mesh element. Simulation of this configuration was took place in a finite element software, Marc.Mentat software. A load of 400 N was applied to the proximal tibia bone in order to simulate stance phase of a gait cycle. From the findings, the pin placement at the second cortex of bone provided optimum stability in terms of stress distribution and displacement, which should be considered for better treatment for transverse tibia fracture. On the other hand, the pin placement at first cortex should be avoided to prevent many complications.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | Finite element analysis, pin placement, external fixator, transverse tibia fracture, stability |
Subjects: | T Technology > TP Chemical technology |
Divisions: | Chemical Engineering |
ID Code: | 84971 |
Deposited By: | Fazli Masari |
Deposited On: | 29 Feb 2020 13:49 |
Last Modified: | 29 Feb 2020 13:49 |
Repository Staff Only: item control page