Universiti Teknologi Malaysia Institutional Repository

An evaluation of single plane ultrasonic tomography sensor to reconstruct three-dimensional profiles in chemical bubble column

Rahiman, M. H. F. and Rahim, R. A. and Rahim, H. A. and Green, R. G. and Zakaria, Z. and Mohamad, E. J. and Muji, S. Z. M. (2016) An evaluation of single plane ultrasonic tomography sensor to reconstruct three-dimensional profiles in chemical bubble column. Sensors and Actuators, A: Physical, 246 . pp. 18-27. ISSN 0924-4247

Full text not available from this repository.

Official URL: https://www.scopus.com/inward/record.uri?eid=2-s2....

Abstract

In this paper we evaluated a single plane ultrasonic tomography sensor and performed some experimentation to reconstruct three-dimensional profiles in a chemical bubble column. An array of 32 ultrasonic sensors was constructed and clamped on the exterior circumference of the column. The information of the profiles was captured based on time-of-flight and arrival-time analysis. Several phantom profiles were created and sampled using ultrasonic tomography techniques and the results were compared with high-speed camera images. The findings showed promising results where the reconstructed three-dimensional profiles matches with the high-speed camera images.

Item Type:Article
Uncontrolled Keywords:Bubble columns, Cameras, High speed cameras, Sensors, Arrival time, High-speed camera image, Three-dimensional profiles, Time of flight, Ultrasonic tomography, Tomography
Subjects:T Technology > TK Electrical engineering. Electronics Nuclear engineering
Divisions:Electrical Engineering
ID Code:72279
Deposited By: Fazli Masari
Deposited On:21 Nov 2017 03:28
Last Modified:21 Nov 2017 03:28

Repository Staff Only: item control page