Dambatta, M. S. and Murni, N. S. and Sudin, Izman and Kurniawan, Denni and Froemming, Gabriele R. A. and Hermawan, Hendra (2015) In vitro degradation and cell viability assessment of Zn-3Mg alloy for biodegradable bone implants. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 229 (5). pp. 335-342. ISSN 0954-4119
Full text not available from this repository.
Official URL: http://dx.doi.org/10.1177/0954411915584962
Abstract
This article reports the in vitro degradation and cytotoxicity assessment of Zn-3Mg alloy developed for biodegradable bone implants. The alloy was prepared using casting, and its microstructure was composed of Mg2Zn11 intermetallic phase distributed within a Zn-rich matrix. The degradation assessment was done using potentiodynamic polarization and electrochemical impedance spectrometry. The cell viability and the function of normal human osteoblast cells were assessed using 3-(4,5-dimethylthiazol-2-yl)-5-(3-carboxymethoxyphenyl)-2-(4-sulfophenyl)-2H-tetrazolium and alkaline phosphatase extracellular enzyme activity assays. The results showed that the degradation rate of the alloy was slower than those of pure Zn and pure Mg due to the formation of a high polarization resistance oxide film. The alloy was cytocompatible with the normal human osteoblast cells at low concentrations (<0.5 mg/mL), and its alkaline phosphatase activity was superior to pure Mg. This assessment suggests that Zn-3Mg alloy has the potential to be developed as a material for biodegradable bone implants, but the toxicity limit must be carefully observed
Item Type: | Article |
---|---|
Uncontrolled Keywords: | osteoblast, Zn-Mg alloy |
Subjects: | T Technology > TJ Mechanical engineering and machinery |
Divisions: | Mechanical Engineering |
ID Code: | 55811 |
Deposited By: | Fazli Masari |
Deposited On: | 06 Oct 2016 06:06 |
Last Modified: | 15 Feb 2017 01:09 |
Repository Staff Only: item control page