Optimization of the paper permeability tester using Robust design

Watanabe, Kazunori and Miyagi, Zenichi and Dolah, Rozzeta and Takahashi, K. (2014) Optimization of the paper permeability tester using Robust design. Jurnal Teknologi (Sciences and Engineering), 68 (4). pp. 91-96. ISSN 0127-9696

[img]
Preview
PDF
900kB

Official URL: http://dx.doi.org/10.11113/jt.v68.3004

Abstract

This paper is using practical case study of robust design engineering. Effect of control factors and optimum condition is studied to design a more robust Paper Permeability Tester. In this study, robustness means decreasing uncertainty that is validity of measurement results with the Tester. Robustness is required to ensure the reliabilityfor the measurement results in the field of the measurement tester. Thus, orthogonal array L18 is employed and the optimum condition was obtained for decreasing the measurement dispersion. Experiment reproducibility of Signal-to-Noise Ratio and sensitivity were confirmed within 20%. SNR value is linearly depends on air permeance value. When the value of air permeance is small, it tends to provide small SNR. Flow rate in nozzle was considered as a major cause. Stability of flow in the nozzle is significant to design the Tester to be more robustness.In a case of a flow under the rapidly, SNR is small due unstable flow. Sensitivity with nozzle diameter was dramatically large. Nozzle specification tolerance wasfound to have strong influence on measurement result

Item Type:Article
Uncontrolled Keywords:permeability tester, permeance, robust parameter design, signal-to-noise ratio, tolerance design
Subjects:T Technology
Divisions:Razak School of Engineering and Advanced Technology
ID Code:54377
Deposited By: Siti Nor Hashidah Zakaria
Deposited On:05 Apr 2016 07:00
Last Modified:12 Aug 2018 03:49

Repository Staff Only: item control page