Abd. Aziz, Azila (2009) Mathematical modeling of an amperometric glucose sensor: the effect of membrane permeability and selectivity on performance. Jurnal Teknologi (Kejuruteraan) (51F). pp. 77-94. ISSN 0127-9696
|
PDF
- Published Version
237kB |
Official URL: http://www.penerbit.utm.my/onlinejournal/51/F/51si...
Abstract
Interference from electro-active chemicals such as ascorbic acid, uric acid and acetaminophen can be a problem for peroxide based glucose biosensors. Most works focused on the employment of a perm-selective membrane sandwiched between the electrode and the active component of the sensor to overcome this problem. In this work, a mathematical model has been developed to study the effect of membrane permeability and selectivity on peroxide based glucose biosensor performance. Digital simulation was carried out using the finite difference method. As expected, membrane selectivity to peroxide played a major role in interference reduction. However, interestingly, the model also suggested that the manipulation of the transport properties of the protective outer layer would also result in acceptable interference reduction.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | Glucose biosensors; mathematical modeling; acetaminophen; interferents; transport properties |
Subjects: | T Technology > TP Chemical technology |
Divisions: | Chemical and Natural Resources Engineering |
ID Code: | 9995 |
Deposited By: | Norhayati Abu Ruddin |
Deposited On: | 16 Jul 2010 04:20 |
Last Modified: | 22 Jul 2010 02:04 |
Repository Staff Only: item control page