Universiti Teknologi Malaysia Institutional Repository

Finite element analysis of mini implant biomechanics on peri-implant bone

Hisam, Muhammad Juzaili and Zaman, Izzuddin and Sharif, Safian and Kurniawan, Denni and M. Nor, Fethma (2019) Finite element analysis of mini implant biomechanics on peri-implant bone. In: 14th Global Congress on Manufacturing and Management, GCMM 2018, 5 - 7 December 2018, Brisbane, Australia.

[img]
Preview
PDF
928kB

Official URL: http://dx.doi.org/10.1016/j.promfg.2019.02.044

Abstract

Mini dental implant whose diameter is between 1.8 and 2.4 mm is a dental implant design currently implemented as bone screw in in orthodontics, as support for denture, and in situations when smaller diameter implant is the feasible option. Biomechanics of the peri-implant bone inserted by mini dental implant is of interest in this study where relevant studies are lacking. This study was intended to investigate using finite element analysis the induced stress and strain on peri-implant bone when a mini dental implant is loaded. The thread pitch of the mini dental implant and the peri-implant bone type were varied, with a constant loading (100 MPa pressure) applied on the mini implant. First, the mini dental implant with three different thread pitches (0.5 mm, 1.0 mm and 1.5 mm) were inserted into a type II bone. It was found that the higher the thread pitch, the higher the maximum stress (increased from 53.2 to 78.6 MPa) and the less distributed the stress on the peri-implant bone. Next is a mini dental implant with 1.0 mm thread pitch was inserted into peri-implant bone types II, III and IV. When the bone type changes from II to III, the maximum stress becomes lower (from 57.8 to be 51.7 MPa) but more high stress was distributed in the cortical bone. The strain was more than doubled (from 0.82 to be 1.76%) on the cancellous bone. When the bone type changes from III to IV, the maximum stress was doubled (from 51.7 to be 104.8 MPa) and more high stress was distributed over the cortical bone. In cancellous bone, the maximum stress was lower (from 9.1 to be 5.3 MPa), but the strain increases almost three folds (from 1.76 to be 5.07%).

Item Type:Conference or Workshop Item (Paper)
Uncontrolled Keywords:Finite element analysis, Mini dental implant, Peri-implant bone, Strain, Stress
Subjects:T Technology > TJ Mechanical engineering and machinery
Divisions:Mechanical Engineering
ID Code:97118
Deposited By: Widya Wahid
Deposited On:23 Sep 2022 01:30
Last Modified:23 Sep 2022 01:30

Repository Staff Only: item control page