Wan Hassan, Wan Muhamad Saridan and Darwish, Z. A. (2011) Comparison of two methods of noise power spectrum determinations of medical radiography systems. In: Malaysia Annual Physics Conference 2010 (PERFIK-2010). American Institute of Physics, New York, pp. 321-323. ISBN 978-073540883-8
Full text not available from this repository.
Official URL: http://dx.doi.org/10.1063/1.3573765
Abstract
Noise in medical images is recognized as an important factor that determines the image quality. Image noise is characterized by noise power spectrum (NPS). We compared two methods of NPS determination namely the methods of Wagner and Dobbins on Lanex Regular TMG screen-film system and Hologic Lorad Selenia full field digital mammography system, with the aim of choosing the better method to use. The methods differ in terms of various parametric choices and algorithm implementations. These parameters include the low pass filtering, low frequency filtering, windowing, smoothing, aperture correction, overlapping of region of interest (ROI), length of fast Fourier transform, ROI size, method of ROI normalization, and slice selection of the NPS. Overall, the two methods agreed to the practical value of noise power spectrum between 10 -3-10-6mm2 over spatial frequency range 0-10mm-1.
Item Type: | Book Section |
---|---|
Uncontrolled Keywords: | full field digital mammography, noise power spectrum, screen film |
Subjects: | G Geography. Anthropology. Recreation > GE Environmental Sciences |
Divisions: | Others |
ID Code: | 28898 |
Deposited By: | Liza Porijo |
Deposited On: | 04 Dec 2012 00:48 |
Last Modified: | 05 Feb 2017 00:09 |